Herunterladen Inhalt Inhalt Diese Seite drucken

Sartorius PMA.Vision VIS1X Installationsanleitung Seite 25

Farbmischwaagen für den einsatz in explosionsgefährdeten bereichen der zone 1
Vorschau ausblenden Andere Handbücher für PMA.Vision VIS1X:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Deutsch
a = Up to three letters and/or numbers or blank (not critical to safety)
b = Up to
two
letters and/or numbers or blank (not critical to safety)
c = Up to five letters and/or numbers or blank (not critical to safety)
Specific Conditions of Use
1. The PMA.Evolution shall only be used with the YCO-14Z
Ex-Link
Converter
2. The protective ground connectors on the PMA.Evolution and the YCO14-Z
Ex-Link
Converter shall be connected
2
together using a potential equalisation conductor with a cross-sectional area of at least 4 mm
.
3. The
CAT6
cable used between the PMA.Evolution and the YCO14-Z
Ex-Link
Converter shall be obtained from
Sartorius.
Equipment Ratings:
YCO14-Z
Ex-Link
Converter
Associated intrinsically safe apparatus for connection to Class I, Division 1, Groups C and D hazardous
(classified) locations in accordance with drawing 2021459, Associated Intrinsically Safe for connection to
Class I, Zone 1, Group IIB hazardous (classified) locations in accordance with drawing 2021459. Ambient
temperature range 0°C to 40°C
PMA.Evolution EVO1Xab-c
PMA.Evolution LAB1Xab-c
Intrinsically Safe for Class I, Division 1, Groups
C and
D, hazardous (classified) locations in accordance
with drawing 2021459, Intrinsically Safe for Class I, Zone 1, Group IIB hazardous (classified) locations in
accordance with drawing 2021459. Temperature class T4 at an ambient temperature of 0°C to 40°C.
FM Approved for:
Sartorius Lab Instruments GmbH &
Co.
KG
Goettingen, Germany
To verify the availability of the Approved product, please refer to www.approvalguide.com
FM Approvals HLC 5/13
«PeopleSoft_Project_ID»
Page 2 of 3
Installationsanleitung VIS1X
25

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis