Technical data

8.1 General

Operating gas:
Required operating pressure:
MMAD (mass median aerodynamic
diameter):
Mass fractions over 5µm:
Minimum fill:
Maximum fill:
The MMAD was determined using a 4% solution of lidocaine hydrochloride
monohydrate in the Next Generation cascade impactor (NGI).
The flow rate in the NGI was 15l/min.
Determination of operating pressure and flow
- The measurement described in the following may only be carried
out by hospital staff, appropriately trained staff at the dealership,
or employees of PARI GmbH.
- The PARI TIA nebuliser S will not function according to specifica-
tion unless the hospital compressed air supply or the compressor
used satisfies the following criteria.
• Connect the nebuliser to a source of compressed air.
• Measure the flow under ambient conditions of +23°C air temperature and
50% relative humidity downstream from the nozzle.
The operating pressure is the pressure present at the nozzle for this flow.
Both values are measured at the same time.
40
Air
1.1bar to 1.7bar
(depending on compressed air
source)
7.3µm to 8.3µm
(depending on compressed air
source)
69.9% to 74.9%
(depending on compressed air
source)
2ml
8ml