Herunterladen Inhalt Inhalt Diese Seite drucken

Miscellaneous; Nozzle Blocked - Pari TIA Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

7

Miscellaneous

7.1 Nozzle blocked

Should the nozzle be blocked, you can clear it yourself as follows:
• Unscrew the nebuliser upper sec-
tion (2) from the nebuliser lower
section (1).
• Push the medication cup (5) up as
far as the windows and then pull it
out of the nebuliser lower section.
• Press the nozzle (7) together with
the airflow control (6) down and out
of the medication cup.
• Pull the airflow control up and off
from the nozzle.
• Take the nozzle cleaner (8) out of
the nebuliser lower section.
• Carefully poke through the three
nozzle bore holes "A" to "C" (only
from below).
• Re-assemble the nebuliser (see
Chapter 4.1 "Assembling the
nebuliser").
- Only use an original PARI nozzle cleaner!
- Use the nozzle cleaner very carefully, because the small nozzle
bore holes can easily be damaged irreparably.
- If the nozzle becomes blocked repeatedly, replace it (part no.
019B2110).
38
2
5
1
C
A
8
6
7
B