Programme 8: Modulation
The frequency is changed continuously within the stated parameter borders.
The pulse width is adapted accordingly.
Minimal frequency: 2 Hz
→ Pulse width: 200 µs
Maximal frequency: 80 Hz
→ Pulse width: 100 µs
Duration of modulation: 7.5 s
Duration of therapy: 30 min
Programme 9: Muscle training
The intensity is altered continuously in accordance with the stated
parameters.
Frequency: 50 Hz
Pulse width: 250 µs
Intensity modulation:
2 s rise → 5 s working → 1 s fall → 12 s pause
Duration of therapy: 30 min
Programme 10: Gate control dynamic 1
The intensity is altered continuously in accordance with the stated parameter,
whereby both channels operate at delayed times to one another (massage).
Frequency: 80 Hz
Pulse width: 150 µs
Intensity modulation: 1 s rise → 1 s fall
Duration of therapy: 30 min