Insufficient blood sample.
The test does
Defective test strip.
not start after
Sample applied after the
applying the
device is automatically
sample.
turned off.
Defective device.
Error in performing
the test.
Control solution vial was
poorly shaken.
The control
Expired or contaminated
solution testing
control solution.
result is out of
range.
Control solution that is too
warm or too cold.
Defective test strip.
Device malfunction.
Specifications
Memory
1000 measurement results with respective date and time
Dimensions
89.8 (L) x 54.9 (W) x 18 (H) mm
Power Source
One 1.5V AAA alkaline battery
Weight
46.1 g (without battery)
External Output
Bluetooth
Auto electrode insertion detection
Auto sample loading detection
Features
Auto reaction time count-down
Auto switch-off after 2 minutes without action
Temperature warning
8°C to 45°C (46.4°F to 113°F), below 85% R.H. (non-
Operating Condition
condensing)
Storage/Transportation
-20°C to 60°C (-4°F to 140°F), 10% to 93% R.H
Condition
Repeat the test using a new test strip
with larger volume of blood sample.
Repeat the test with a new test strip.
Repeat the test with a new test strip.
Apply sample only when flashing "
appears on the display.
Please contact customer service.
Read instructions thoroughly and
repeat the test again.
Shake the control solution vigorously
and repeat the test again.
Check the expiration date of the
control solution.
Control solution, device, and test
strips should be at room temperature
(20°C to 25°C / 68°F to 77°F) before
testing.
Repeat the test with a new test strip.
Please contact customer service.
Blood Glucose Test: mg/dL or mmol/L
Hematocrit Test: %
Hemoglobin Test: g/dL
Measurement Units
β-ketone Test: mmol/L
Total cholesterol Test: mg/dL
"
Uric acid Test: mg/dL
Blood Glucose Test: 10 ~ 600 mg/dL (0.55 ~ 33.3 mmol/L)
Hematocrit Test: 0 ~ 70%
Hemoglobin Test: 0 ~ 23.8 g/dL
Measurement Range
β-ketone Test: 0.1 ~ 8.0 mmol/L
Total cholesterol Test: 100 ~ 400 mg/dL (2.6 ~ 10.4 mmol/L)
Uric acid Test: 3 ~ 20 mg/dL (0.179 ~ 1.190 mmol/L)
Blood Glucose Test: 0 ~ 70%
Blood Glucose/Hematocrit/Hemoglobin Test: 0 ~ 70%
Hematocrit Range
β-ketone Test: 10 ~ 70%
Uric Acid Test: 20 ~ 60%
Total Cholesterol Test: 20 ~ 60%
Blood Glucose Test: Capillary/ Venous/ Neonatal/ Arterial
Blood Glucose/Hematocrit/Hemoglobin Test: Capillary/
Venous/ Neonatal/ Arterial
Test Sample
β-ketone Test: Capillary/ Venous
Uric Acid Test: Capillary
Total Cholesterol Test: Capillary
Test Result
Measurements are reported as plasma equivalents
This device has been tested to meet the electrical and safety requirements of: IEC/
EN 61010-1, IEC/EN 61010-2-101, EN 61326-, IEC/EN 61326-2-6, EN 301 489-17,
EN 300 328.
EN-9