β-Ketone Test:
Message
< 0.1 mmol/L
> 8.0 mmol/L
Total Cholesterol Test:
Message
< 100 mg/dL (2.65 mmol/L)
> 400 mg/dL (10.4 mmol/L)
Uric Acid Test:
Message
< 3 mg/dL (0.179 mmol/L)
> 20 mg/dL (1.190 mmol/L)
Error Message
Error
Cause
Message
There is insufficient power
E-b
in the batteries to perform
the test.
The strip has been used.
The blood sample has been
E-U
applied to the strip before
the meter was ready for the
measurement.
What it Means
What it Means
What it Means
What To Do
Replace the batteries immediately
and reset the date and time on the
meter setting.
Repeat the test with a new strip.
Repeat the test with a new strip.
Apply the blood sample when the
meter indicates that it is ready for the
measurement by displaying "
and a "
" which starts blinking.
E-E
Meter calibration data error.
The component of the device
E-0
might be broken.
Meter malfunction due to
E-R
damaged components.
Incorrect data on the meter or
E-C
the code strip.
You may have removed the
strip after applying blood,
E-F
or there is insufficient blood
volume.
Ambient temperature is below
the operating temperature.
E-t
Ambient temperature is above
the operating temperature.
Appears when the test strips
are expired. (This only applies
E-2
to β-ketone, total cholesterol
and uric acid test strips)
Blood Glucose Measurement
Symptom
Batteries exhausted.
The device does
not display a
message after
Test strip inserted upside
inserting a test
down or incompletely.
strip.
Defective device or
"
test strips.
EN-8
Review the instructions and repeat
the test with a new strip and using
the correct technique. If the problem
persists, contact your local customer
service for assistance.
Review the instructions and repeat the
test with a new test strip. Please make
sure you use the correct technique and
dose with enough blood.
The operating temperature range is
from 8°C to 45°C (46.4°F to 113°F).
Repeat the test after the device and
test strip have reached the above-
mentioned operating temperature.
Ensure the date and time settings of
the meter are correct and up to date.
Insert the code strip again. If the
problem persists, please check the
expiry date of the test strip.
Discard the expired test strips. Use
a new test strip that is in date and
repeat the test.
Cause
What To Do
Replace the battery immediately
and reset date and time on the
meter setting.
Insert the test strip with contact bars
end first and facing up.
Please contact customer service.