Troubleshooting
8
Troubleshooting
Personnel:
Protective equipment:
Fault description
Unit does not function
'Container reserve'
lights up in display
'Container empty' lights up in
display and red LED flashes
'Alarm' and 'Rotorspray' light up in
display and red LED flashes
'Alarm' and 'Solenoid valve' light
up in display and red LED flashes
When the cause of the problem has been removed, press 'ENTER' to
acknowledge the fault. When the cause of the problem has been removed
the applicator is ready to work again.
417101357 Rev. 4-02.2020
n
Specialist
n
Qualified electrician
n
Mechanic
n
Service personnel
n
Chemical resistant protective gloves
n
Protective gloves, mechanical hazards
n
Safety boots
n
Protective goggles
Cause
Wrong mains voltage
Defective fuse
Reserve signal active
Supply container is
empty
Cable has been pulled
out or is defective
although container is
full
Rotor spray not
connected but active in
parameter mode
Rotor spray connection
cable defective
Rotor spray defective
Connection cable to
solenoid valve
defective
Solenoid valve
defective
30
Remedy
Check the applicator adjustment
(see
Ä Chapter 6.1.1.1 'Setting
the mains voltage' on page 24)
Replace fuse
Fill container (the unit is ready for
operation even if it is not refilled
Ä Chapter 9.1 'Changing
(see
the application tip' on page 31)
Ä Chapter
Refill product (see
9.1 'Changing the application tip'
on page 31 )
Connect or replace empty signal
cable
Connect rotor spray or make rotor
spray inactive in parameter mode
Ä Chapter 7.4 'Parameter
(see
settings (see flow diagram)'
on page 28)
Replace connection cable
Replace rotor spray
Replace connection cable
Replace solenoid valve