The tester switch "OFF" completely when the rotary switch is in
the "OFF" position.
The tester has a "Hold the last reading before stopping the test".
The tester is fuse protected and has a crowbar between C
C
. This crowbar is activated by voltage. If the voltage is too
2
high, that crowbar will blow the fuse automatically to interrupt the
circuit.
The voltage between P
voltage, but does not have a fuse.
The tester has a temperature shut down. The temperature
sensing is done on a current regulation transistor. Should this
over-temperature LED lit, allow the instrument to cool down for a
while before proceeding further.
4. Specifications
Electrical
Measuring ranges
Accuracy
Test current
Test current accuracy
and P
is also protected for over
1
2
0-200.0 m in steps of 100 µ
0-2000 m in stepf of 1m
0-20.00 in steps of 10 m
0-200.0 in steps of 100 m
0-2000 in steps of 1
+/- 0,5% of reading +/- 2 digits
over the operating temperature
range,
supplied test leads
1 mA in 2000 range
10 mA in 200/20 range
100 mA in 2000 m/200 m range
+/- 0,5%
-20-
-15°C to +55°C, with the
and
1