6. Faults

6. Faults
Faults
Status display shows red
warning dot on the dial
EN
The pointer position is
arbitrary and does not
correspond to the process
pressure.
Medium can ingress into
the reference chamber.
Despite a damaged
diaphragm element, due
to the second barrier no
exchange has occurred
between process and
ambience
Following any failure or shut-down of the plant all instruments should
be checked and, if necessary, replaced before recommissioning the
plant.
14
Causes
Measuring
instrument is
defective and
can no longer
be used as a
reference for
the process
pressure.
WIKA operating instructions pressure gauge, model PG43SA-D
Measures
Replace measuring
instrument. The
dismounting of the
measuring instru-
ment is described
in chapter 8.