5. Commissioning, operation
5.6 Zero adjustment
External zero adjustment, setting range ±15°
5.7 Autoclaving
Observe the permissible conditions of autoclaving, see chapter
9.3 "Permissible temperature"
During the autoclaving process, no weight-loading must be exert-
ed via the case and the instrument must not lie on its window.
After the first autoclaving cycles, the window slightly arches
outward. This has no effect on the function of the instrument.
Depending on the autoclaving process, whitish residues can build
up on the window and pointer. This has no effect on the function of
the instrument.
In case of a zero offset, carry out a zero adjustment. See chapter
5.6 "Zero adjustment".
WIKA operating instructions pressure gauge, model PG43SA-D
The zero adjustment can be made
with a slotted screwdriver after the
removal of the sealing plug on the
top of the case.
Deviations in the display due to
natural fluctuations in the atmos-
pheric pressure and altitude varia-
tions (up to 2,000 m height above
sea level) can be compensated
with the zero adjustment.
EN