h) Sorting mode

Sorting mode permits quick selection of components according to the measured reference value, an individually
adjustable reference value and specified tolerance areas. The parameter settings are changed with the arrow but-
tons (6).
Proceed as follows to set the sorting mode:
Connect the reference component to the measuring input (11) or the measuring lines.
Press the "SORTING" button (5) to activate sorting. The display shows "Sorting" and the measured value is saved
as a reference. The pre-set tolerance value is ±1%. When the tolerance specification meets your criteria, continue
with the sorting measurement. The test result is displayed in the main display as "PASS" (measured value in the
tolerance range) or "FAIL" (measured value outside of the tolerance range). The subdisplay shows the measured
value.
If you want to manually enter the reference value or change the tolerance, proceed as follows:
Push the button "SETUP" (7) to enter the settings for the measuring range, the reference value and the tolerance.
The settings are performed in sequence.
You are in the menu item "Set measuring range". The icon "RANGE" flashes in the display. Use the two arrows left/
right (6) to select the measuring range.
Confirm the selection with the "ENTER" (7) button.
You are in the menu item "Set reference value". The smallest digit flashes in the display. Use the arrows buttons
up/down (6) to change the value. Use the two arrows left/right (6) to select the decimal digit. You may enter a value
from 1 to 1999.9 counts. Confirm the selection with the "ENTER" (7) button.
You are in the menu item "Set tolerance range". The current tolerance value flashes in the display. Use the two
arrows left/right (9) and (11) to select the tolerance range. You have the following options: ±0.1% ±0.2% ±0,25%
±0,5% ±1% ±2% ±5% ±10% ±20% and -20% to +80%. Confirm the selection with the "ENTER" (15) button.
You may continue the sorting measurement. The test result is displayed in the main display as "PASS" (measured
value in the tolerance range) or "FAIL" (measured value outside of the tolerance range). The subdisplay shows the
measured value.
Push "SORTING" (5) to terminate sorting.

i) uSB-interface

A mini-USB port built-in on the top side of the measuring instrument can be used to transfer measurement data to a
PC for further processing.
Data connection can be established by connecting the cable included in the scope of supply with a free USB port at
your PC.
For this purpose, insert the interface cable into the USB port (13) on the upper side of the housing.