Herunterladen Diese Seite drucken

Minebea Intec Midrics 1 Ex Bedienungsanleitung Seite 121

Werbung

Hazardous s Location n
Class s I, , II, , III, , Div v 1, , GP P A,B,C,D,E,F,G
Class s I, , Zone e 1, , Grp. . IIC
Class s II, , Zone e 21
Class s III
-10°C C (+14°F) ) < < T
< < +40°C C (+104°F)
AMB
RS232
RS422
Interf ace
Interf ace
Terminal:
Terminal:
J800
J801
U
(V
) = 12.6V*
i
ma x
U
(V
) = 25.2V**
U
(V
) = 8.6Vdc
i
ma x
i
ma x
I
(I
) = 328mA***
I
(I
) = 175mA
i
ma x
i
ma x
P
(P
) = any
P
(P
) = 500mW
i
ma x
i
ma x
C
= 550pF
C
= 221nF
i
i
L
= 0mH
L
= 0mH
i
i
U
(V
) = 6Vdc
U
(V
) = 6Vdc
o
oc
o
oc
I
(I
) = 50.5mA***
I
(I
) = 30.3mA
o
sc
o
sc
P
= 37.87mW
P
= 37.87mW
o
o
C
(C
) = 39.9µF
C
(C
) = 39.7µF
o
a
o
a
L
(L
) = 13.92mH
L
(L
) = 38.7mH
o
a
o
a
L
(L
)/R
= 395µH/
L
(L
)/R
=800.26µH/
o
a
o
o
a
o
[20] / [21]
[20] / [21]
[19]
alternative connection
Note:
*: versus ground
**: between the lines
***: resistiv ely limited
21.11.2022
T. Hiller
Indicator: MIXS1-...
DMS
UNI-IN
Interf ace
Terminal:
Te rminals:
J102
J400
U
(V
) = 8.61Vdc
U
(V
o
oc
o
oc
I
(I
) = 213.5mA
I
(I
) = 86.2mA
o
sc
o
sc
P
= 459.53mW
P
= 351mW
o
o
C
(C
) = 4.65µF
C
(C
) = 526nF
o
a
o
a
L
(L
) = 777.47µH
L
(L
) = 4.78mH
o
a
o
a
L
(L
)/R
= 77.37µH/
L
(L
)/R
o
a
o
o
a
[22]
[23]
Passive device
(e.g. switch)
[22]
[19]
Any NRTL approved load cell or
weighing platform with Entity
Concept parameters.
e.g. CAPX... or IUX.4-... or IFX.4-...
Any NRTL approved equipment with
Entity Concept parameters.
e.g.: MIXS1-... (2nd display mode)
Supply
Terminals:
J100, J101
U
(V
) = 12.6Vdc
i
ma x
I
(I
) = 205mA
i
ma x
P
(P
) = 2.6W
i
ma x
C
= 2.44nF
i
L
= 0µH
i
) = 8.61Vdc
U
(V
) = 6Vdc
o
oc
I
(I
) = 20.2mA
o
sc
P
= 25.25mW
o
C
(C
) = 39.9µF
o
a
L
(L
) = 86.8mH
o
a
=101µH/
L
(L
)/R
= 1407µH/
o
o
a
o
[24]
[24]
[19]
[28]
[9]
[9]
Control Drawing
67002-751-07 (EN)
[9]
Hazardous s Location n
Class s I, , II, , III, , Div v 2, , GP P
A,B,C,D,E,F,G
Class s I, , Zone e 2, , Grp. . IIC
Class s II, , Zone e 22
Class s III
Any NRTL approved power
supply with Entity Concept
[25]
parameters.
[18]
e.g.: YPS08-Y...
[9]
Non-Hazardous Location
Any NRTL approved power
supply with Entity Concept
[26]
parameters.
[18]
e.g.: YPS08-Z...
[9]
Any NRTL Approved Associated
equipment with entity concept
[27]
parameters.
e.g.: YDI05-Z...
[9]
MIXS1-...
08
Revision
Power
[7]
Power
[7]
Data
1
4
of
Sheet

Werbung

loading

Diese Anleitung auch für:

Mixs1