Follow the prompts on the screen of the control unit. For those filters
t
where it is not possible to scan the barcode on either the filter itself or
the packaging, manual entry is provided in the Ambr
ware.
5.4

Removing the Filter for Clean-in-Place

Requirements
The operator is prompted to remove 1 or more filters from the Ambr
flow UF | DF Module on the screen of the control unit. This should be done
before running the clean-in-place (CIP) with the CIP fitting provided.
Procedure
NOTICE Leaks due to loose compression fittings!
t
When connecting | disconnecting tubing on the product, make sure
t
to only handle the Luer Lock ring part (1) of the connector.
Do not twist the threaded compression fittings (2).
t
If the threaded compression fitting gets accidentally undone, screw it
t
back on by hand and tighten as much as possible.
Unscrew the Luer Lock rings (1) on the Luer fittings.
t
To disengage, twist and pull the Luer tapers from the filter, ensuring a
t
secure hold on the Luer fitting.
Do not disconnect the tubing from the tube interface block.
t
Once all 3 fittings are disengaged, unclip the hollow fiber filter from the
t
tube clips or if using a flat sheet filter, undo the thumb screws enough to
release the filter clamps and remove the filter.
Remove the filter and discard appropriately.
t