If the filter clamps are horizontally positioned (for flat sheet filters), and
t
you switch to hollow fiber filters, unscrew the thumb screws on each side
of the filter clamps at the top and bottom of the holder.
Turn the filter clamps 90° and reinsert them back in the holder before
t
tightening the thumb screws to secure them in place.
Ensure the bottom clamp is at the top of its slot.
t
The position of the top clamp is adjusted depending on which filter is fit-
y
ted.
Once the filter clamps are placed in the desired orientation on the hold-
t
er, hook the holder onto the adapter.
Optional: Tilt the holder slightly to ensure the 2 prongs (1) at the bottom
t
fit either side of the handle of the filter holder.
When using a flat sheet filter, place the filter in the holder.
t
Adjust the lower clamp as needed.
t
Ensure the barcode (1) is on the opposite side of the tube interface block
t
for future scanning.