Electromagnetic Compatibility Guidance
Electromagnetic Emissions
Table 1 Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The TrueCPR coaching device, model 80596-000001, is intended for use in the
electromagnetic environment specified below. The customer or user should ensure that
the TrueCPR device is used in such an environment.
Emissions Test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC 61000-3-2
Voltage fluctuations/
Not applicable
flicker emissions
IEC 61000-3-3
The TrueCPR device emits a continuous wave signal at 12 KHz. The maximum Effective
Isotropic Radiated Power (EIRP) and Effective Radiated Power (ERP) levels are -40 dBm,
measured at a distance of 3 m (10 ft).
Essential Performance
The TrueCPR device maintains safe and effective performance (accurate depth and recoil)
when operated in the electromagnetic environment specified in Table 2 through Table 4 .
Electromagnetic Immunity
Table 2 Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The TrueCPR device, model 80596-000001, is intended for use in the electromagnetic
environment specified below. The customer or user should ensure that the TrueCPR
device is used in such an environment.
IEC 60601
Immunity Test
Test Level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
©2012 Physio-Control, Inc.
All manuals and user guides at all-guides.com
Electromagnetic Environment - Guidance
The TrueCPR device uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
The TrueCPR device is suitable for use in all
establishments including domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
The TrueCPR device is battery-powered.
Compliance
Level
±6 kV contact
±8 kV air
Electromagnetic Environment -
Guidance
Floors should be wood,
concrete, or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
73