Table 2 Guidance and Manufacturer's Declaration - Electromagnetic Immunity
(Continued)
The TrueCPR device, model 80596-000001, is intended for use in the electromagnetic
environment specified below. The customer or user should ensure that the TrueCPR
device is used in such an environment.
IEC 60601
Immunity Test
Test Level
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV line(s) to
line(s)
IEC 61000-4-5
±2 kV line(s) to
earth
Voltage dips,
<5% U
short
(>95% dip in U
interruptions
for 0.5 cycle
and voltage
40% U
variations on
(60% dip in U
power supply
for 5 cycles
input lines
70% U
IEC 61000-4-11
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 sec
Power
3 A/m
frequency (50/
60 Hz) magnetic
field
IEC 61000-4-8
Note: U
is the AC Mains voltage prior to application of the test level.
T
74
All manuals and user guides at all-guides.com
Compliance
Level
Not applicable
Not applicable
Not applicable
T
)
T
T
)
T
T
)
T
T
)
T
3 A/m
Electromagnetic Environment -
Guidance
The TrueCPR device is battery-
powered.
The TrueCPR device is battery-
powered.
The TrueCPR device is battery-
powered.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
TrueCPR Device Instructions for Use