Event Review Screens
Two event review screens display data for the most recent device use. To display
the event review screens, start with the device turned off. Press the
and
(POWER) buttons simultaneously to turn the device on in Event Review
mode. To switch between the screens, press the
Troubleshooting Tips
These troubleshooting tips provide immediate corrective actions for possible
problems. If a problem persists, contact your local Physio-Control representative
for assistance. In the USA, call Technical Support at 800.442.1142.
OBSERVATION
No audible metronome.
Device does not function.
Display is visible, but does not
provide compression feedback.
Alert indicator persists.
©2012 Physio-Control, Inc.
All manuals and user guides at all-guides.com
Percent of compressions with depth in target
range.
Percent of compressions with proper recoil.
Average compression rate (compressions per
minute).
Total event duration.
Percent of total event with compressions.
CORRECTIVE ACTION
• Use compression rate meter for guidance.
• Only use device within specified operating
temperature range of 0°–40°C (32°–104°F).
• Perform CPR unaided.
• Turn device off and back on. Do not apply
pressure to chest pad until calibration is
complete and feedback screen appears.
• Perform CPR unaided.
• Turn device off and back on. If indicator
persists, perform CPR unaided.
• Contact Technical Support or your local
Physio-Control representative for
assistance.
English
(AIRWAY)
(MUTE) button.
11