5.5. Discontinuing a measurement

If it is necessary to interrupt a blood pressure measurement for any
reason (e.g. the patient feels unwell), the START button can be
pressed at any time.
The device immediately lowers the cuff pressure automatically.
5.6. Memory - recall of measurements
This blood-pressure monitor automatically stores the last
30 measurement values. By pressing the «M» button, the
last Measurement, as well as previous measurements can
be displayed one after the other.
mmHg
mmHg
MR
(MR30: Value of the last measurement)
5.7. Memory– cancellation of all measurements
Attention!
Before you delete all readings stored in the memory, make
sure you will not need refer to the readings at a later date.
In order to delete all stored readings, depress the «M»
button for at least 7 seconds, the display will show the
symbol «CL» and 3 short beep sounds will be heard to
indicate deletion of stored readings.
6.

Error Messages/Malfunctions

If an error occurs during a measurement, the measurement is
discontinued and a corresponding error code is displayed.
(Example: Error no. 1)
Error No.
Possible cause(s)
ERR 1
The systolic pressure was determined but afterwards the cuff pressure fell below 20 mmHg.
The tube may have become unplugged after the systolic blood pressure was measured.
Further possible cause: No pulse has been detected.
ERR 2
Unnatural pressure impulses influence the measurement result. Reason: The arm was moved
during the measurement (Artefact).
8
mmHg
mmHg
MR
mmHg
mmHg
MR
(MR29: Value of the measurement before MR 30)
M
mmHg
mmHg
MR
M