Problem
Inaccurate
dispensing
Inaccurate
dispensing with
certain liquids
Not dispensing
18 English
Possible cause
Incorrect operation
Tip incorrectly attached
Calibration altered: caused by
misuse, for example
Unsuitable calibration. High viscosity
liquids may require recalibration
Pistons stuck or not connected
Solution
Follow the instructions carefully.
Attach firmly
Recalibrate according to the
instructions.
Recalibrate with the liquids in
question.
Remove the tip cone module. Move
the piston by hand or with piston
removal tool. Attach the module in
service mode.