Herunterladen Diese Seite drucken

Hach TenSette plus Benutzerhandbuch Seite 12

Vorschau ausblenden Andere Handbücher für TenSette plus:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5. Fill a clean reagent reservoir with the liquid to be dispensed.
6. Dip the tip under the surface of the liquid in the reservoir and press the trigger. This action will fill
the tip and the first volume is displayed.
7. Withdraw the tip from the liquid touching it against the edge of the reservoir to remove excess
liquid.
8. Deliver the first volume by pressing the trigger. The display shows the next volume.
9. Continue dispensing by repeating step 3. After the last step, Blowout is displayed.
10. To empty the tip completely, press the trigger.
11. Change the tip and continue pipetting.
Note: The pipette can be emptied at any time by pressing Cancel (left selection key, refer to
on page 8).
6.6.4 Mix + Dilute
1. Select Program > Edit, then select with the scroll key Mix + Dilute and accept with OK.
2. Select the pipetting first volume by pressing the scroll key up or down.
3. Accept the volume with OK.
4. Select the second volume (VOL 2) with the scroll key and accept with OK.
5. Select the pipetting speeds.
6. Fill a clean reagent reservoir with the liquid to be dispensed.
7. Dip the tip under the surface of the first liquid in the reservoir and press the trigger. The first
volume is drawn in to the tip. Air is displayed.
8. Withdraw the tip from the liquid touching it against the edge of the reservoir to remove excess
liquid. Press the trigger again to aspirate the air buffer.
9. Dip the tip under the surface of the second liquid and press the trigger. The second volume is
drawn in to the tip.
10. Withdraw the tip from the liquid.
11. To deliver the both volumes, press the trigger. Mix is displayed.
12. By pressing the trigger the pipette starts to pipette approximately 70% of the total volume several
times as long as the trigger is held pressed.
13. After releasing the trigger, the pipette stops after the next dispensing and Blowout is displayed.
14. Withdraw the tip from the liquid and press the trigger and hold it down to empty the tip.
15. Release the trigger to return to the ready position.
16. Change the tip and continue pipetting.
6.6.5 Manual
With manual mode it is possible to measure volumes. In manual mode only slower speeds are
available to ensure a quick stop.
1. Select Program > Edit, then select with the scroll key Manual and accept with OK.
2. Then select a limit for the total volume, the default is the maximum volume.
3. Select the pipetting speeds. 0.00 is then displayed.
4. Fill a clean reagent reservoir with the liquid to be dispensed.
5. Dip the tip under the surface of the liquid in the reservoir and press the trigger. The liquid is drawn
in to the tip and the actual volume is shown on the display.
6. The pipetting direction can be changed with the In/Out key (left selection key, refer to
on page 8).
7. To deliver the liquid, select the Out key (down direction) and press the trigger.
8. To deliver the rest of the liquid, press the trigger and hold it down and hold it down to empty the
tip completely.
12 English
Figure 3
Figure 3

Werbung

loading

Diese Anleitung auch für:

Bbp087