Instruction Manual PCR Workstation Pro
Resetting the operating hours of the UV Air Recirculator UV tube
After renewing the UV Air Recirculator UV tube the operating hours should be set back to zero.
To reset the time
Push << Mode >> 5 times until you see the operating hours screen.
•
Push << Set >>. You will see the Reset Surface Screen.
•
To reset the operating hours push << + >>.
•
To go back without resetting the operating hours push << - >>.
•
[Warning] Exposure to UV light is harmful.
The polycarbonate front panel is UV blocking and will not allow UV radiation to
pass through the panel. The UV lamp will shut off if the front panel is opened.
The PCR Workstation Pro contains a powerful source of UV radiation that will
cause damage to unprotected eyes and skin. Before operating any unit, be sure
all personnel in the area are properly protected. Even though the unit shuts the UV off when the
front panel is open, UV blocking eyewear should be worn as well.
Extended periods of time UV exposure can be used to decontaminate the interior chamber by inac-
tivation of DNA. Refer to "Measuring the UV tube intensity", page 18, for information on measuring
the intensity of the UV tubes with a UV meter.