Instruction Manual PCR Workstation Pro

Replacing the UV tube of the UV Air Recirculator

Locate and remove the six screws on the ventilation cover
on the outer back panel of the workstation.
Pull away the ventilation cover from the workstation.
The tube has two prongs on each end that fit into a socket,
then rotates into place.
Carefully hold the tube on the ends and insert the tube into
the socket. Twist the tube 90° until the tube locks into
place.
Replace the cover back into position and reattach the six
screws.

Replacing the fuse

The fuse is located at the back right side of the system. To remove
the fuse, use a flathead screwdriver to open the lid and pull out the
fuse.
Insert the new fuse and using the screwdriver, turn the fuse clock-
wise to lock.

Replacing the filter of the UV Air Recirculator

The fan filter of the UV Air Recirculator is placed on the in-
ner top of the PCR Workstation Pro in front of the light tubes
(see picture to the right). To change the filter, take off the
plastic cover with the filter attached to it and replace it with a
new filter. Plunge the plastic cover back into its fixation.
PEQLAB_v1116_E