Replacing the UV Air Recirculator UV tube

Locate and remove the six screws on the ventilation cover on
the outer back panel of the workstation.
Pull away the ventilation cover from the workstation.
The tube has two prongs on each end that fit into a socket, then
rotate into place.
Carefully hold the tube at the ends and insert the tube into the
socket. Twist the tube 90° until the tube locks into place.
Replace the cover back into position and reattach the six
screws.
We recommend you change the UV tubes after 1000 – 1500
hours of operation.

Replacing the HEPA/UV tube

Turn the unit power off and unplug the unit from the power
outlet. Locate and remove the four screws on the pre-filter
cover on the left side of the workstation.
Pull away the pre-filter and open the security cover by removing
the one screw.
The tube has two prongs on each end that fit into a socket, then
rotate into place.
Carefully hold the tube at the ends and insert the tube into the
socket. Twist the tube 90° until the tube locks into place.
Replace the security cover and the pre-filter back into position
and reattach the four screws.
We recommend you change the UV tubes after 1000 – 1500 hours
of operation.

Replacing the filter of the UV Air Recirculator

The UV Air Recirculator fan filter is located on the inner top of
the PCR Workstation in front of the light tubes (see picture to the
right). To change the filter, take off the plastic cover with the filter
attached to it and replace it with a new filter. Push the plastic
cover back into its location.

Replacing the HEPA and Carbon filter

Turn the unit power off and unplug the unit from the power
outlet. Lift the front panel and locate the HEPA filter cassette.
The filter cassette has two handles on the left and the right
side of the cassette. By pressing and pulling them downward
the cassette will be removed and the Carbon and HEPA filters
can be replaced.
To load the filter cassette, unpack the filter from foil and add
the HEPA filter first (arrow downwards, for correct direction of
flow) followed by the carbon filter.To place the filter cassette
into the top assembly move it upwards into the opening and
press gentle until the handles locks into place.