Herunterladen Inhalt Inhalt Diese Seite drucken

Setting Up A Daily Decontamination Routine - VWR PEQLAB PCR Workstation Pro Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für PEQLAB PCR Workstation Pro:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Instruction Manual PCR Workstation Pro
To run a shorter decontamination press << Mode >>.
The screen will show "Decontaminated". Time can be changed from 5 to 30 min in 5 min
steps by pushing the << + >> and << – >> buttons.
Push << Set >> to set the default UV light decontamination time to the indicated value.

Setting up a daily decontamination routine

To set up a daily start program that starts a decontamination program every day at the
same time push << Mode >> two times until the display shows "Daily Start".
Push << Set >> to change the settings.
By pushing << + >> the program can be turned "ON".
To confirm push << Set >>.
The cursor will automatically switch to set up the minutes.
Change minutes by pushing << + >> and << - >>.
Push << Set >> to confirm.
The cursor will jump to hours.
Change the hour to turn on the PCR Workstation Pro by pushing << + >> and << - >>.
Push << Set >> to confirm.
The PCR Workstation Pro will now start a decontamination program every day at the set time (e.g.
for 30 minutes at 10:30)
PEQLAB_v1116_E
15

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis