Refit all mechanical components in reverse order to removing them. Return the
power supply to the unit.
2.2
How to make an software up-date and set-up
In order to run the 0.8 bar pressure line sampling option the unit must be fitted with a
special software. This update should be done either by a service technician from
Endress+Hauser or a skilled and fully trained person.
The requirements to enable a software up-date are a lap top PC with the ReadWin
2000 operating programme and an original diskette containing the special
programme.
Special software (LDB400a) for the ASP Station 2000 water sampler for
pressurised pipe lines
General:
The water sampler fulfils all the requirements of the standard product. All basic
information such as technical data, mechanical and electrical installation,
commissioning, set-up, safety instructions etc. can be found in the operating manual
for the standard product.
All technical changes and expansions to the standard product are contained and
explained in these additional operating instructions.
15mm
New fixing holes:
∅ 4,5mm
20mm
15