Wiring
RTD
x2
x5
x6
4-W
3-W
2-W
TC
Typ J,K,T,N... ±100mV...±1V
R12 R11 R13 R22 R21 R23
L|+ N|-
 7
Terminal assignment of the device (channel 2 and relays as an option)
If long signal cables with high-energy transients can be expected, we recommend fitting
a suitable surge arrester upstream.
5.1.1
Overview of possible connection options at the device
34
I
x4
x4
x8
x7
x8
0/4...20mA
U<=1V
x7
x8
x7
Terminal assignment of analog inputs, channel 1 and 2 (optional)
11
12
13
15 16 17 18
LPS 2-W
x1
x2
x4
x8
4...20mA
U>1V
x3
x8
x8
0...10V
0/1...5V
±10V
±30V
Analog Out
4...20mA
14
21
22
23
25 26 27 28
LPS 4-W
x1
x2
x4
x8
0/4...20mA
21
22
23
24
25 26 27 28
11
12
13
14
15 16 17 18
24
RIA45
A0010228
A0010406
Endress+Hauser

Werbung

loading