Herunterladen Inhalt Inhalt Diese Seite drucken

Calibrating The Measuring Device - VOLTCRAFT SDM-2125 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Deleting statistical data
1
You can delete the statistical data (such as the measurement number (NUM), the average
(AVG), the minimum values (MIN) and the maximum values (MAX)).
Press and hold the Clear button
data are reset to zero. Measuring can start again.

13. Calibrating the measuring device

Press and hold the confirmation button (4) to switch to the desired calibration mode (zero
calibration
The calibration mode is set as outlined above in section 'e) Setting calibration mode' in the
section 'Menu settings'. Zero calibration and two-point calibration are available.
Carrying out zero calibration
1
If you have selected zero calibration
If the following figure is shown on the LC display (2), place the sensor (8) directly on the
uncoated calibration disc. Wait for approx. 2 seconds before you lift the measuring device
off again.
1
After lifting, a zero value is displayed. Calibration is complete.
Carrying out two-point calibration
1
If you have selected two-point calibration
the non-metallic layer thickness standard (of known thickness) over it.
If the following image is shown on the LC display (2), place the sensor (8) directly on the
calibration disc with the standard layer thickness plate. Wait for approx. 2 seconds before
you lift the measuring device off again.
(5) for two seconds to delete all saved data. All statistical
or two-point calibration
).
, use an uncoated calibration disc for calibration.
2
2
, use an uncoated calibration disc and place
47

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

2239527

Inhaltsverzeichnis