13.2 Stability Detection Band

(up to 8 counts for verifiable, up to 64 counts for non verifiable models)
The band makes it possible to select conditions at which the balance should be
considered as stable. If „1 count" is selected, and the reading (within counting the
readings) remains constant, the balance is considered as stable and the
display is shown. The stability detection band can be set from 2 to 64 counts.
Select the menu element:
27
28
29
30

13.3 Tracking

Tracking is the function which ensures to receive present value displaying in possible
long period.
To ACTIVATE this function, select the menu element 34.
To DEACTIVATE this function, select the menu element 35.

14 Capacity display

This function displays a bar graph representation of the load on the weighing pan.
This may be used to prevent sudden appearance of "oL" (overload) during
measurement.
Select the menu element 11 in the menu overview to set the full range mode:
(1)
(2)
If the bar graph is not to be displayed, press the menu element 21
PBS/PBJ-BA-def-2319
for
The bar located within the lower scale range indicates
that the weight located on the scale pan is small. (1)
The bar approaching the upper scale range indicates
that the weight located on the scale pan almost
reaches the balance capacity. (2)
stability
1 count
2 counts
4 counts
8 counts
52