8.2.1 Adjustment test with external weight

• Condition: Menu setting „E-tESt" / Menu element 4
• At verified balances, the adjustment test is locked by a switch (except
accuracy class I). To disable the access lock, destroy the seal and actuate
the adjustment switch. 9Position of the adjustment switch see chap. 9
Attention:
After destruction of the seal the balance must be re-verified by an authorised
agency and a new verification wire/seal mark fitted before it can be reused
for applications subject to verification.
 In weighing mode press
If „E-tESt" is not displayed, press
activate menu element 4, see menu overview.
 Press
, checking is started. The displayed
weight value of the recommended adjustment weight
(see chapter 1) flashes.
If the weight value shall be changed, press
desired setting using navigation buttons (see chapter 3.1.1 „Numeric input").
 Carefully place adjusting weight in the centre of the
weighing plate ,press
 Wait until the zero display flashes.
PBS/PBJ-BA-def-2319
. E-tESt is displayed.
to return into weighing mode and
, the active digit flashes.
.
Carry out the