Remove adjustment weight and press
successful adjustment the balance automatically returns to
weighing mode.
In case of an adjustment error (e.g. objects on the weighing
plate) the display will show an error message, repeat
adjustment.
When an optional printer is connected and the GLP function
is connected, the adjustment log will be edited, see chap.
8.5. Printout example (KERN YKB-01N):
PBS/PBJ-BA-def-2319
- - - - - - - - - - - - - - - - -
CAL –EXTERNAL
- - - - - - - - - - - - - - - - -
KERN
TYPE PBS4200-2M
SN WBxxxxxxxxx
ID 1234
DATE 27-01-2011
TIME 11.54.53
REF 4000.00g
BFR 3999.97g
AFT 4000.00g
-COMPLETE
-SIGNATURE-
- - - - - - - - - - - - - - - - -
. After
Adjustment mode
Company
Model
Serial no.
Balance identification no. (see chap.
8.5.1)
Used adjustment weight
Before adjustment
After adjustment
prepared by
28