8.2.2 Adjustment test with internal weight

Condition: Menu setting „I-tESt"/ menu element 2
 In weighing mode press
If „i-tESt" is not displayed, press
activate menu element 2, see menu overview.
 Press
, checking is automatic.
Wait for a short period of time, the difference to the previous
adjustment is displayed.
 Either
Press
, the d-value is reset to zero.
Adjustment of balance by resetting.
or
 Press
, the d-value is not reset to zero. No
adjustment occurs.
PBS/PBJ-BA-def-2319
. I-tESt is displayed.
to return into weighing mode and
32