Error message:
Displayed error
code
* Please contact your sales representative.
PBS/PBJ-BA-def-2319
Explication
Disturbances in weight loading
mechanical elements.
Weight on the scale pan is unstable
during calibration.
Large zero point drift during
calibration.
Large drift during the PCAL function.
Large drift during calibration of the
measuring range.
Calibration weight is incorrect.
Disturbances in the balance (the
balance is stopped when this display
appears).
Received command code is
incorrect.
Total number of the displayed unit is
longer than 7 positions.
Disturbance in the balance.
The power voltage is incorrect.
Remedy
Check transport screws.
Empty the scale pan.
Use the correct weight.
Use the correct weight.
Use the correct weight.
Check limiters etc.
Reduce load.
Check the power voltage .