Power Failure

In the event of a power failure (momentary or long term), the Turbo-
V ICE controller will stop the turbopump and all the interconnected
pumps/devices. The Turbo-V ICE vent valve device, if used, will vent
the turbopump only if the power failure is longer than the preset
delay time. When power is restored, the Turbo-V ICE controller
automatically restarts the interconnected devices and the turbopump
in the proper sequence.

Remote Control Mode Operation

If remote signals are used to operate the controller, it must be
programmed for remote operation (see paragraph " Operating
parameter selections " ) and when ready to start, the display shows:
R
S
If the Soft Start has been deselected the display shows:
P
With or without Soft Start mode selected the START/STOP and LOW
SPEED front panel push- buttons are inoperative, while the CYCLE
NUMBER and PUMP CURRENT pushbuttons are always active.
Turbo-V 1000 ICE Controller User Manual / 87-900-905-01 (M)