Turbo-V 1000 ICE Controller User Manual / 87-900-905-01 (M)
DESCRIPTION
START S2 signal 24 V, 60 mA, optically isolated output (pin 12 positive, pin 4
negative).
The output Voltage will be present when the START push-button on front panel
is pressed or the remote start is present. or the function has been requested by
RS 232/422/485, until NORMAL operation is reached.
OVERTEMP. signal 24 V, 60 mA, optically isolated output (pin 13 positive pin 5
negative).
The output Voltage will be present when a fault condition is displayed on the
front panel display, with only in an overtemp condition
FAULT signal 24 V, 60 mA, optically isolated output (pin 13 positive pin 6
negative).
The output Voltage will be present when a fault condition is displayed on the
front panel display, with exception of an overtemp condition
Analog output voltage (0 - 10 V) of pump speed (pin 7 positive, pin 14 negative).
Analog output Voltage of DC current drawn by the turbopump (pin 8 positive,
pin 15 negative). 1 Vdc proportional to 1 A.
Flow meter alarm, the flow value remained below the threshold for 10 seconds
or more.
15
9
Typical output connection open collector
Technical Information
8
8
15
+ 24
9
1
1
17
Interconnections
+24 V