Herunterladen Inhalt Inhalt Diese Seite drucken

Window-Based Protocol - Agilent Technologies Turbo-V 2000HT 969-9562 Bedienungshandbuch

Vorschau ausblenden Andere Handbücher für Turbo-V 2000HT 969-9562:
Inhaltsverzeichnis

Werbung

Window-Based Protocol

The following table, valid for the RS 232, RS 422 and RS 485
interfaces, describes each single window used in the protocol.
Tab. 7
WIN
000
001
100
101
102
103
104
106
107
108
109
200
201
202
203
205
206
207
208
209
210
Turbo-V 2000HT Controller User Manual / 87-900-913-01 (O)
R
W
T
Description
X
X
L
START/STOP
X
X
L
LOW SPEED [0=OFF / 1=ON]
X
X
L
SOFT START [0=NO / 1=YES]
X
X
L
DEAT TIME
X
X
L
PROCESS GAS [0=Argon / 1=Air] Default Argon
X
X
N
SPEED THRESHOLD
X
X
N
RUN UP TIME (0÷359.999 sec)
X
X
N
HIGH SPEED ADJUST
X
X
N
MODE (0, 1, 2) [FRONT, REMOTE, SERIAL]
X
X
N
BAUD_RATE (0-4) [600, 1200, 2400, 4800, 9600]
X
L
PUMP LIFE RESET (TYPE "ON" TO RESET)
X
N
CURRENT [ A]
X
N
VOLTAGE [ V]
X
N
POWER [ W]
X
N
FREQUENCY [ Krpm]
X
N
PUMP STATE (0÷6) [STOP, WAITING INTERLOCK,
STARTING, NORMAL, HIGH LOAD, FAILURE,
APPROACHING]
X
N
ERROR CODE (0÷7) [NO ERROR, OVERVOLTAGE,
SHORT CIRCUIT,CHECK CONN.,
X
L
STATE K1 [0=OFF / 1=ON]
X
L
STATE K2 [0=OFF / 1=ON]
X
N
UPPER BEARING TEMPERATURE (0÷90) [ °C]
X
N
LOWER BEARING TEMPERATURE (0÷90) [ °C]
Technical Information
Optional Serial Port
[0=NO / 1=YES]
17
263/306

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis