PROGRAMMING THE LM OG-AM AND LM OG-TAERM

LM OG-A (CND)

The units of measure and the correction factor can be configured in the programming mode . A program button on the
backside of the register will enable the mode, only accessible when the register has been removed from the meter . Removing
the register from the meter will break the sealing . This seal has to be restamped by the national authorities (MID) .
01
Turn on the register by pressing the
TOTAL button .
02
Press the programming button on the
backside of the register to enter the
programming mode .
03
The measuring unit will flash and can be
changed by pressing the RESET button
(L, GAL, QT, PT) .
To move forward, press the TOTAL button .
04
Press TOTAL to change the k-factor . The
number to be changed is flashing and can
be modified with RESET .
To change the next number, press TOTAL .
05
The second number can now be changed .
To move to the next number, press
TOTAL again .
06
The third number can now be changed .
To move to the next number, press
TOTAL again .
07
The fourth number can now be changed .
To move to the next number, press
TOTAL again .
Seite / Page / Page 22
AUT-UM-03446-DE-02
08
The fifth number can now be changed .
To move forward, press TOTAL again .
09
The arrow for the flow direction flashes
now . Use RESET to change or TOTAL to
move forward .
10
Press the programming button on the
backside of the register to quit the
programming mode .
11
The register goes into the sleep mode .
12
To wake up the register, press the
RESET button .
13
Press RESET for 3 seconds and the
register displays the checksum and serial
number in the bottom line .