Herunterladen Inhalt Inhalt Diese Seite drucken

Switch From Totalizer (L) To Flow Rate (L/Min); Change Pulse Rate; 6-Digit Totalizer With Two Decimal Places - Badger Meter LM OG Serie Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

LM OG Baureihe / LM OG Series / Série LM OG

Switch from Totalizer (l) to Flow Rate (l/min)

Change Pulse Rate

Predefined pulse rates:
1/2" = 100 ppl
3/4" = 61 .5 ppl
1" = 61 .5 ppl
1-1/2" = 17 ppl

6-digit Totalizer with Two Decimal Places

Seite / Page / Page 18
Press and hold RESET button .
Then press TOTAL .
To switch back the programming to the totalizer, press RESET .
Press TOTAL 3 times then press RESET 3 times (quickly) .
You are now in programming mode .
Then press and hold TOTAL .
The figure to be changed is flashing and can be modified with RESET .
To change the next figure press TOTAL .
Press TOTAL + RESET simultaneously to save the programming .
Currently programmed pulse rate: 100 .00 PPL
An additional 6-digit totalizer with 2 decimal places can now be called up in the register .
Press and hold TOTAL .
Current Totalizer: 0 .04L
AUT-UM-03446-DE-02
April/April/Avril 2023

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis