PRODUCT MAINTENANCE
Fig. 27:DK50 2x2V/110
230 V
23.9. Procedure for connecting a compressor to a new cabinet
Prior to any maintenance or repair work, switch off the compressor and
disconnect it from the mains (pull out the mains plug).
The compressor in a cabinet requires that the jumper is not mounted in the terminal strip for
correct operation (Fig. 26, Fig. 27 pos. A). The switch on the cabinet switches the entire product
on and off, including the compressor.
If there is a jumper in the terminal strip of the compressor connected to a
cabinet, the switch on the compressor cabinet will not work!
If we want to connect a new cabinet to a compressor that previously worked alone (a compressor
without a cabinet), then it is necessary to cancel the connection in the compressor terminal strip
by performing the procedure, see further. (Fig. 26, Fig. 27). (See also chap. 13.1 and 13.2).
PROCESS:
Jumper removal (D-C-B-A process)
 Disconnect the product from the mains by pulling the plug out of the socket.
 Remove the cover on the electrical panel (on the compressor).
 The jumper is in the terminal strip – D
 Lift up the fuse cover – C
 Remove the jumper from the terminal strip – B
 Close the fuse cover – A
 Reinstall the cover on the electrical panel.