Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Data; Order List - Dräger Alcotest 9510 Standard IR Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

12

Technical data

Ambient conditions
Operational characteristics
Temperature range
Humidity
Ambient pressure
Operating altitude
Storage
Temperature range
Humidity
Ambient pressure
Measuring system
Measuring range
Breath alcohol concentration 0 to 3 mg/l
Blood alcohol concentration 0 to 6 ‰
Resolution final result
Measurement duration
(typical)
Blowing time
Respiratory volume
Repeatability:
0 to 0.4 mg/l
above 0.4 to 2.0 mg/l
above 2.0 mg/l
Calibration interval
Typical operating data
Supply voltage
Mains operation
Dräger Alcotest 9510 Standard IR / EC
0 to 40 °C
30 to 90 % relative humidity
850 to 1060 hPa
max. 3000 m
–20 to 70 °C (>40 °C only for
some hours)
10 to 90 % relative humidity
700 to 1300 hPa
0 to 300 µg/100 ml
0 to 3000 µg/l
(Ethanol mass per breath
volume at 34 °C and
1013 hPa.)
0 to 6 g/l
0 to 0.6 %
(ethanol mass per blood
volume at 20 °C and
1013 hPa. When giving this
data a conversion factor of
2100 between breath and
blood alcohol concentration is
used.)
0,01 mg/l
1 µg/100 ml
10 µg/l
0,01 ‰
0,01 g/l
0,001 %
3 minutes
>3 seconds
>1.5 l
Standard deviation
< 0.007 mg/l
Less than 1.75 % relative
Less than 6 % relative
Depends on required accuracy
Recommended: 6 months;
max.: 12 months
110 to 240 V, 50 to 60 Hz
12 V operation
Power consumption
During the
warming-up phase
in standby mode
Main fuse
Thermal printer
Time indication
Warm-up phase
Dimensions
Weight (without keypad)
CE markings
13

Order list

Designation
Dräger Alcotest 9510 Standard IR
Dräger Alcotest 9510 Standard IR / EC
Consumables and accessories
Keypad - Mini QWERTY (USB)
Carrying bag
12 V cable (with ON/OFF switch)
Printer paper (5 rolls, storable for 7 years)
Printer paper (5 rolls, storable for 25 years)
Mouthpiece with non-return valve (25 pieces)
Fuse, TH 250, 0.63 A
Stylus (3 off)
Mains cable (certified according to EN 60950)
Removable cover
Screen protection foils
Technical data
10.5 to 15.5 V DC
approx. 70 W
<15 W
TH/250 V; 0,63 A, 2 off
Thermal printer with special
paper, 58 mm wide
Day/Month/Year
Hours: Minute
Summer time/winter time
less than 15 minutes at room
temperature
approx. 325 mm x 250 mm x
55/185 mm (W x H x D)
6,9 kg
Electromagnetic compatibility
(Directive 89/336/EEC)
Low Voltage Directive
(Directive 72/23/EEC)
Order no.
83 19 460
83 19 470
83 19 441
83 19 155
83 12 641
83 19 002
83 18 461
68 05 703
18 92 979
83 18 402
83 24 658
83 18 479
83 20 999
27

Quicklinks ausblenden:

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Dräger alcotest 9510 standard ec

Inhaltsverzeichnis