Making measurement

4
Making measurement
4.1

Requirements

Measurements of breath alcohol concentration with the
Alcotest 9510 may only be carried out by persons trained on
the instrument.
Ensure that the permissible ambient conditions (see Section
12 on Page 27) are met.
Waiting and control time before measurement
Basically, the following has to be considered for breath alcohol
measurement:
Observe a waiting time of 20 minutes after the last
consumption of alcoholic beverages before starting the
measurement.
It must be ensured that during a time period of at least
10 minutes before the measurement no substances are
taken in through the mouth or nose (control time).
Alcohol residues that are still in the mouth may cause
incorrect measuring values.
This may also occur by aromatic drinks (e.g. fruit juices),
alcoholic mouth sprays, medicines and drops, as well as by
burping and vomiting.
Rinsing the mouth with water or non-alcoholic beverages
does not replace the waiting or control time!
4.2

Starting measurement

1. Press the start button to trigger the measuring process.
If the instrument is in energy saving mode (STANDBY is
displayed), press the start button again after the warming-
up phase to start the measurement.
Dependent on the instrument configuration the test subject
data are entered prior to the measurement.
The start of the measurement is indicated by the message
PREPARING TEST.
Complete each entry of corresponding data with Next or
Reset.
4.2.1

Entering data of person being tested

The figure shows the entry via
the virtual keyboard. When
using an external keypad, the
entries
must
be
made
correspondingly
via
keypad. With  it is possible to
change between small and
capital letters.
NOTICE
i
i
If no entry is made within 10 minutes, the message
DATA INPUT TIMEOUT is displayed and the
instrument returns to READY.
Up to 24 letters can be entered for a name. For names
with more letters/characters the last character The
entry is concluded with Next.
20
Alcotest 9510
LUEBECK DD.MM.YYYY
OPERATOR'S FAMILY NAME
1
2
3
4
5
6
7
8
Q
W
E
R
T
Y
U
I
this
A
S
D
F
G
H
J
K
Z
X
C
V
B
N
M
Reset
Prev.
Next
LUEBECK
The entry prompt NAME OPERATOR is displayed.
1. Enter the operator name and confirm with Next.
The entry prompt SUBJECT'S FAMILY NAME is displayed.
2. Enter the test subject name and confirm with Next.
The entry prompt SUBJECT'S FIRST NAME is displayed.
3. Enter the first name of the test subject and confirm with
Next.
The entry prompt SUBJECT'S DATE OF BIRTH is displayed.
4. Enter the digits for the date of birth (DD), month of birth
(MM) and year of birth (YYYY) and confirm with Done.
NOTICE
i
i
If implausible dates are entered, data acceptance is
refused after selecting the button Summary and a
fresh date entry is requested.
Up to this time all previous entries can be recalled on
the display with Prev.. The entries can either be
modified by new entries or confirmed with Done.
A summary of the data entered
is displayed. By double click on
the individual entries these can
be edited again.
5. Confirm the entry of test subject data with Next. The
measuring
confirmation. Scrolling back through or modifying previous
entries no longer possible.
NOTICE
i
i
If the external keyboard is being used, the data entry
can also be completed by entering the key
combination [ alt ] + [ s ].
The measuring process is continued.
PURGING is displayed.
The measuring system is purged with ambient air. Next the
internal reference point is determined, BLANK CHECK is
displayed.
hh:mm
Only in the Alcotest 9510 Standard IR / EC this is followed
by a test of the ambient air for alcoholic vapours using the
9
0
EC sensor, ENVIRONMENT CHECK is displayed.
O
P
:
If interference from the ambient air can be precluded during
L
,
+
this test, the device prompts for providing a breath sample.
-
The time displayed in the bottom section indicates the time
at which provision of the breath sample must be started.
Done
6. Open the packaging of a new mouthpiece, take the breath
04323875_en.eps
hose out of the storage tray and insert the mouthpiece
without skin contact into the handle at the end of the breath
hose by using the packaging.
Alcotest 9510
OPERATOR
Klein
SUBJECT'S FAMILY NAME
Smith
SUBJECT'S FIRST NAME
Paul
SUBJECT'S DATE OF BIRTH
18.09.1956
Reset
process
starts
immediately
Dräger Alcotest 9510 Standard IR / EC
hh:mm
LUEBECK DD.MM.YYYY
Save
LUEBECK
02523875_en.eps
after
the

Quicklinks ausblenden:

Werbung

Kapitel

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Dräger alcotest 9510 standard ec

Inhaltsverzeichnis