System Configuration
A pharmaceutical water ozonation system (shown in Figure
1 below) typically uses three ozone measurement points:
1.
Following the ozonated water storage tank.
2.
Following the UV ozone destruction lamp to confirm
no ozone is going to the distribution loop.
3.
At the end of the distribution loop, to verify that
adequate levels of ozone have reached the entire
distribution loop during sanitization with the UV
ozone destruction lamp off.
Point 3
Ozone
sensor
Vent
Storage
tank
Figure 1: Typical pharmaceutical ozone loop
A semiconductor ultrapure water system typically does not
shut down for sanitization and omits the third measurement
point.
pureO
3
Dissolved Ozone Sensor
pureO
Ozone generator
PID control
Point 1
Ozone
sensor
© 07 / 16 Mettler-Toledo GmbH
Dissolved Ozone Sensor
3
Points of use
UV
Point 2
Ozone
sensor
Printed in Switzerland