Herunterladen Inhalt Inhalt Diese Seite drucken

Parameter Factory Settings - HBM DIS2116 Kurzanleitung

Waagenelektronik
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

DIS2116
11

Parameter factory settings

When the DIS2116 is delivered, it comes with specific factory settings:
• The (7-digit) production number is entered
As the DIS2116 has undergone certain tests during production, the following
information can deviate from zero:
• Legal-for-trade counter (Menu "INFORMATION SCALE" or F1 key with
symbol "i")
• Sensor-overflow counter (Menu "INFORMATION BUSSCAN")
Parameter
Limit values 1...2
Filter mode
Filter
Manual tare value
COM1, baud rate
COM1, parity bit
COM2,COM3,COM4, Function
COM2,COM3,COM4,baud rate
COM2,COM3,COM4,parity bit
Printing, ESC sequences
Basic scale function
Menu access
Legal-for-trade operation
Unit
Zero on start-up
Zero tracking
Standstill monitoring
Increment
Decimal point
Maximum capacity
Multi-range scale
Calibration weight
Gravitational acceleration
correction
Scale char. curve, zero
Scale char. curve, nom. value
Linearization
When restoring the factory settings ("FACTORY SETTINGS DIS2116" menu),
the parameters described above are set to their default values.
A2540-1.1 en/de
Default value
0
5
0
38400
even
OFF
9600
even
0
Standard
5
OFF
OFF
OFF
OFF
OFF
1d
0
10000
0
10000
9,8102
0
10000
Explanation
Deactivated
Standard filter
0.3 Hz (AD104C, C16i)
Deactivated
Deactivated
Deactivated
Non-automatic scale
All menus
Deactivated
No unit of measurement
Deactivated
Deactivated
Deactivated
right
Deactivated
Max. capacity adjustment
Deactivated
At load cell maximum capacity
Deactivated
35