7.2 Switching on and off

The device is switched on once the supply voltage is applied.
When the electronics are switched on, a BUS-SCAN is executed initially, i.e. a
check is made as to whether the load cells listed in the PARAMETER MENU
− SCALE CONFIGURATION are present.
During initial commissioning, all digital load cells have the same address 31
(factory setting). The load cells must be configured, before a scale adjustment
can be carried out.
With verified scales, a check is made as to whether the parameters in the load
cells have been changed. If the parameters in the load cells are no longer
identical to those during verification, an error message is displayed and no
measured values.
Connected load cells that are not entered in the "PARAMETER MENU →
SCALE CONFIGURATION" are not taken into account.
During the BUS-SCAN:
• Information line 1 displays an identification string consisting of the manu-
facturer, TYPE, serial number and software version
• Information line 2 displays date and time
• Measured value status line displays the legal-for-trade counter status
(TCR) and whether the scale is legal for trade (LFT)
Zero is automatically set during initialization if this function is activated.
The actions carried out during initialization are displayed in clear text in the
measured value display and the initialization progress is shown by a moving
bar.
The scale must not be loaded before it is switched on.
The device is switched off by disconnecting the supply voltage from DIS2116.