Basic scale functions

All device functions can be controlled in one or more of the following ways:
• Front control with 19 short-stroke keys. With the function keys F1 ... F4, the
meaning of the keys is displayed in the display function line.
• An external keyboard which can be connected via the PS2 input
• Connecting an external computer via the COM2 serial interface
The main scale functions (Gross/Net, Taring, Zero setting) are controlled via
the function keys F1...F4. For calibration and further device settings a menu is
called up with the key
not interrupted during parameter input, right up to exiting the menu. When
computer commands are used for control, measurement generally continues
without interruption. The exceptions to this are settling after filter selection and
power failsafe storage to the EEPROM.

7.1 Zero setting the scale

Press the F4 key until the symbol >0< appears above the F3 key.
Press the F3 key to set the measurement value to zero.
The displayed gross value is zero immediately after zero setting.
MODE ³
LEGAL-FOR-TRADE
NO (not leg. f. trade)
OIML, NTEP
The % figures relate to the nominal weighing range (parameter "SCALE PA-
RAMETERS → PARAMETER → NOMINAL VALUE")
Execution is dependent on standstill recognition (Chapter 7.9, page 25).
The net display is deactivated.
NOTE
Zero setting is only possible when the function in the "PARAMETER MENU →
FUNCTION KEYS → ZERO SETTING" is set to ACTIVE.
A2540-1.1 en/de
(or F5 on the external keyboard) . Scale operation is
Range of zero setting, lower limit
− 20 %
− 2 %
Range of zero setting, upper limit
+ 20 %
+ 2 %
21