Verfügbare Sprachen

  • DEUTSCH, seite 153
Voltage dips
0% U
0% U
and
Voltage
70% U
interruptions
cycles
0% U
IEC
cycles
61000-4-11
Rated power
frequency
30 A/m
magnetic
50 or 60 Hz
fields
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
B EN PIIS DFU WEB / Rev.04 / 02-2018 (Old ZF 1902127.EN)
; 0.5 cycle
0% U
T
; 1 cycle
0% U
T
and
; 25/30
70% U
T
cycles
; 250/300
0% U
T
cycles
30 A/m
50 or 60 Hz
Mains power quality
should be that of a typical
public low-voltage power
; 0.5 cycle
T
supply network that
supplies buildings used
for domestic purposes,
; 1 cycle
T
commercial or hospital,
clinical environment. If
; 25/30
the user of the
T
ProPex
battery charging during
power mains
interruptions it is
; 250/300
T
recommended that the
ProPex
powered from a separate
power supply (UPS, etc.).
Power frequency
magnetic fields should be
at levels characteristic of
a typical public low-
voltage power supply
network that supplies
buildings used of
domestic purposes,
commercial or hospital,
clinic environment.
®
II requires
®
II charger be
47/50