Precautions
For apex localization, concentrations of NaOCl higher than 5% may
result in reduced accuracy.
As a safety precaution in order to avoid over-instrumentation, it is
recommended to proceed as follows: place the file onto an
endodontic ruler at the point where the ProPex
'APEX'. Subtract a minimum of 0.5 mm from the measured file
length.
Please also respect the following precautions:
- for your own safety, be aware of wearing personal protective
equipment (gloves, glasses, mask);
- if the bar graph makes sudden large movements in the coronal
part of the canal, slowly continue advancing the file toward the
apex until the signal returns to normal;
- this ProPex
combination with any other apparatus or system. It must not be
used as an integral component of any other apparatus or system.
Using replacement parts or accessories not supplied by the
original manufacturer or vender could adversely affect the EMC
performance of the ProPex
- the device must be used with the manufacturer's original
accessories only;
- unplug the device before replacing the battery;
- never use batteries that are leaky, deformed, discolored or
otherwise abnormal;
- in case of battery leakage, carefully dry the battery terminals and
remove all of the leaked liquid. Then replace the battery with a
new one;
- if you do not use the device for a long period of time, it is
recommended to remove the battery from the device to prevent
possible fluid leakage;
- dispose of old batteries according to local codes and regulations;
- accessories including lip clips, hooks or forks should be clean and
without residue of chemical disinfectants or other medicinal
solutions such as sodium hypochlorite or formalin;
- do not expose ProPex
- ProPex
humidity conditions.
B EN PIIS DFU WEB / Rev.04 / 02-2018 (Old ZF 1902127.EN)
®
II unit must not be connected to or used in
®
II to any liquid;
®
II must be stored in normal temperature (< 60°C) and
®
II;
®
II indicates
9/50