Cleaning, Disinfection and sterilization
Operation
3.
Manual Cleaning
26/50
Operating mode
1. Place the pre-cleaned
accessories into the cleaning
bath for the prescribed contact
®
time (e.g. CIDEZYME
,
®
ENZOL
Enzymatic Detergent
Solutions, Johnson & Johnson
Medical, 0.8% for 1 minute);
make sure that the products are
fully immersed (if necessary, use
a soft brush to carefully brush
them down).
2. Remove the accessories from
the cleaning bath and rinse them
off thoroughly at least three
times for a minute each time with
sterile, deionized water.
3. Next, place the accessories in
an ultrasonic bath with a
cleaning agent (e.g.
®
®
CIDEZYME
, ENZOL
Enzymatic Detergent Solutions,
Johnson & Johnson Medical,
0.8% for 20 minutes).
B EN PIIS DFU WEB / Rev.04 / 02-2018 (Old ZF 1902127.EN)
Warning
When choosing cleaning agents
and disinfectants, make sure
that:
- they are suitable for cleaning or
disinfecting instruments;
- you use a disinfectant with
proven efficacy (e.g. with VAH/
DGHM or FDA certification or
CE mark) and that the
disinfectant is compatible with
the cleaning agent;
- the chemicals used are
compatible with the accessories.
Combined cleaning/disinfection
products should only be used if
the instruments are only slightly
contaminated (no visible
contamination/remnants).
Comply with the concentrations
and contact times specified by
the manufacturers of the
cleaning agents and
disinfectants, as well as their
instructions regarding the
intensity of subsequent rinsing.
Use only freshly prepared
solutions, water that is sterile or
has a low microbe content (<
10 cfu/ml) and a low endotoxin
content (< 0.25 EU/ml, e.g.
purified water (PW/HPW)), and
filtered, oil-free air for drying.
Make sure that the accessories
are not in direct contact with one
another.
The hook mechanism has to be
activated during the cleaning
process, the rinsing process and
the ultrasonic bath cleaning
(press several times the push
button) to allow the inner parts to
be cleaned more effectively.
Make sure that the products are
fully immersed (if necessary, use
a soft brush to carefully brush
them down).
No visible impurities should be
observed on the accessories