Cleaning, Disinfection and sterilization
Operation
4.
Disinfection
5.
Rising
6.
Inspection
B EN PIIS DFU WEB / Rev.04 / 02-2018 (Old ZF 1902127.EN)
Operating mode
1. Once the accessories have
been cleaned and inspected,
place them into the disinfection
bath for the prescribed contact
time (e.g. Cidex OPA, Johnson
& Johnson Medical, 100% for
20 minutes); the accessories
must be sufficiently immersed in
the solution.
2. Remove the accessories from
the disinfection bath and rinse
them off thoroughly at least five
times for a minute each time with
water.
3. Dry the accessories by
blowing them down fully.
Abundant rinsing (at least
1 min).
Inspect devices and sort out
those with defects.
Warning
The hook mechanism has to be
activated several times during
disinfection and the rinse to
allow the inner parts to be
disinfected more effectively.
To dry use oil-free, filtered
compressed air and then leave
the accessories to dry further in
a clean place for at least
20 minutes.
Once the accessories are dry,
inspect and pack them as soon
as possible.
Use quality water in accordance
with local regulations.
If a disinfecting solution contains
a corrosion inhibitor, it is
recommended to rinse the
instruments just before the
autoclaving.
Dry on a single use non- weaved
cloth, or with a drying machine or
filtered compressed air.
Defects include:
- deformation of the plastic;
- corrosion;
- discoloration of the plastic.
Dirty instruments must be
cleaned and disinfected again.
Maintenance is not required.
Do not use instrument lubricant.
27/50