Herunterladen Inhalt Inhalt Diese Seite drucken

Setting Up Model Cpu6000-M - WIKA CPU6000-W Betriebsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6. Commissioning, operation

6.2.3 Setting up model CPU6000-M

1. Connect the CPU6000-M to the PC/laptop using the supplied USB cable.
2. Directly download the driver from the local website and install.
http://en-co.wika.de/cpu6000_w_cpu6000_s_cpu6000_m_en_co.WIKA
→ Software → Interface driver
The driver has been correctly installed when, in Windows Device
Manager, by switching the CPU on or off, a COM port appears under
"Connections".
3. In WIKA-CAL, within "Object manager" under "Workplace", select the new COM
port under the "COM port" of the defined CPU6000-M (multimeter) and save the
data record. The multimeter must first be generated and the work station assigned
(see instruction manual WIKA-CAL, chapter 6.3.17 "Multimeter").
4. Set baud rate 9600 and, via the [Test] button, check the communication. The
communication is working properly if a realistic value is displayed.
WIKA operating instructions CalibratorUnit, model CPU6000