4.2
Maintenance
Depending on the quality of the sample gas it might be required to replace the valves from time to time. The
replacement is described in chapter 6 (appendices).
If the valves are very dirty, especially after a short operation time, filtering upstream the pump should be
considered. This will increase the service time substantially.
5
Maintenance and Disposal
If the device shows irregularities see chapter 6 for troubleshooting
If you need help or more information
call +49(0)2102-498955 or your local agent.
If the device doesn't work correctly after elimination of failures and turning power on, the device must be
checked by the manufacturer. Please ship the device with suitable packing to
Bühler Technologies GmbH
- Service -
Harkortstraße 29
40880 Ratingen
Germany
6
Appendices
6.1
Troubleshooting
Problem / Failure
Pump doesn't start
No flow
Pump is noisy
6.2

Spare parts

Attached you will find spare part drawings for the pump where the names and part numbers are given
Possible cause
No power
valves defective or dirty
bypass valve fully opened
bellow is broken
Crank mechanism worn
BX420001, 10/2006
Check connection, fuse and
switch
Blow out valves carefully or re-
place them
Close bypass valve
Replace bellow
Replace crank mechanism
Art. Nr. 90 31 052
Solution
21